The RAD-60 is a Personal Alarming Radiation Dosimeter. It’s a precise and reliable instrument for ensuring the safety of personnel. Ideally, the RAD-60 is used in stand-alone conditions for everyday radiation monitoring . If your needs grow for a more sophisticated system, the versatile RAD-60 can also be integrated into an Access Control System. The RAD-60 can also be switched into System Mode, for the purpose of tracking Personnel Dose records and generating compliance reports.
The design includes state-of-the-art technology with built in memory for retrieving dose, even during power-down. It eliminates outside interference from shock and RF. The RAD-60 is easily programmed by the user, has a digital display, and operates with a single AAA alkaline battery.
With the push of a button, you can turn the unit on/off, change the digital display to read dose or dose rate, select from several dose and dose rate alarm levels, turn the chirp function on/off, reset the integrated dose, and perform battery tests. The large digital display gives instant dose or dose rate readings with a loud audible alarm.
Energy compensated Si-Diode
Dose: 1 uSv – 9.99 Sv or 0.1 mrem – 999 rem
Hp(10), 60 keV – 3 MeV, better than + or – 25%, up to 6 MeV, better than + or – 35%
5 uSv/h – 3 Sv/h or 0.5 mrem/h – 300 rem/h
Better than + or – 15% up to 3 Sv/h (300 rem/h)
Better than + or -5% (Cs-137, 662 keV at 2 mSv/h), Hp(10)
ACS Express Software, ADR-1 Configuration Kit
1 Year Limited
Front panel button has the following functions:
Seven separate alarms, sound level typically better than 85 dBA at 30 cm
Six preset values each for integrated dose and dose rate-push button selection
Infrared communication via bottom of the dosimeter
One triple A alkaline cell, typical life is 1800 hours in background (dose mode)
-20 – + 50° C operational, humidity up to 90% RH, non-condensed -20 – + 70° C storing
80 g (including battery)
78 x 67 x 22 mm